Arsenic

596

Arsenic

Totals: (Complete Digestion) As

Summary

This method measures the arsenic concentration in plant tissue using digestion with nitric, perchloric and sulfuric acids, reduction of arsenate to arsenite, and determination by Vapor Generation Inductively-Coupled Plasma Emission Spectrometer (VG-ICP). The method has a detection limit of approximately 0.05 ppm.

Sample amount requested: 3 g (for reporting on a 100% dry basis)

Questions concerning limited sample size can be answered by the UC Davis Analytical Laboratory.

Tracy, M. L. and Moeller, G. 1990. Continuous flow vapor generation for inductively coupled argon plasma spectrometric analysis. Part 2: Arsenic. J. Assoc. Off. Anal. Chem. 74:516-521.