Selenium

894

Selenium

Totals: Se

Summary

This method is quantitative for selenium. Samples are digested using perchloric and sulfuric acids and reduced using hydrochloric acid (selenate to selenite (IV)). Determination is by vapor generation Inductively-Coupled Plasma Emission Spectrometer (ICP-AES). The method has a detection limit of 0.5 ug/L.

Sample amount requested: 40 mL

Questions concerning limited sample size can be answered by the UC Davis Analytical Laboratory.

Tracy, M. L. and Moeller, G. 1990. Continuous flow vapor generation for inductively coupled argon plasma spectrometric analysis. Part 1: Selenium. J. Assoc. Off. Anal. Chem. 73:404-410.